Class II

Medical Device Recall: Syva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mL

Siemens Healthcare Diagnostics, Inc. · July 13, 2026

Reason for Recall

Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.

Distribution

Worldwide - US Nationwide distribution in the state of CA and the country of Italy.

States Affected

NATIONWIDE

Quantity Affected

20 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2975-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.