Class II
Medical Device Recall: ARTIS icono ceiling. Model Number: 11328100.
Siemens Medical Solutions USA, Inc · June 8, 2026
Reason for Recall
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
Distribution
US Nationwide distribution in the states of CA, GA, IA, IN, LA, MO, NC, NY, PA, SC.
States Affected
NATIONWIDE
Quantity Affected
57 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2648-2026
Status: ongoing