Class II

Medical Device Recall: ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Proc

Siemens Medical Solutions USA, Inc · May 15, 2025

Reason for Recall

In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.

Product Description

ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the ARTIS family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e., patient extremities. This does not include projection radiography. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. ARTIS can also support the acquisition of position triggered imaging for spatial data synthesis.

Distribution

Worldwide.

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2019-2025

Status: ongoing