Class II

Medical Device Recall: Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

SIN SISTEMA DE IMPLANTE NACIONAL S.A · September 1, 2024

Reason for Recall

Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.

Distribution

US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.

States Affected

CA, CO, FL, ID, IL, KY, NJ, NY, OH, TX

Quantity Affected

123 (US 98, OUS 25)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0920-2025

Status: ongoing