Class I

Medical Device Recall: CADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232 21-2120-0100-233 21-2120-0100-50 21-2120-0100-51 21-2120-0100-95 21

Smiths Medical ASD, Inc. · April 10, 2025

Reason for Recall

Affected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.

Product Description

CADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232 21-2120-0100-233 21-2120-0100-50 21-2120-0100-51 21-2120-0100-95 21-2120-0102-02 21-2120-0102-03 21-2120-0102-06 21-2120-0102-07 21-2120-0102-08 21-2120-0102-12 21-2120-0102-13 21-2120-0102-14 21-2120-0102-15 21-2120-0102-17 21-2120-0102-237 21-2120-0102-238 21-2120-0102-247 21-2120-0102-249 21-2120-0102-51 21-2120-0102-78 21-2120-0102-92 21-2120-0102-97 21-2120-0102-98 21-2120-0103-01 21-2120-0103-02 21-2120-0103-03 21-2120-0103-06 21-2120-0103-07 21-2120-0103-08 21-2120-0103-12 21-2120-0103-14 21-2120-0103-15 21-2120-0103-17 21-2120-0103-249 21-2120-0103-51 21-2120-0103-78 21-2120-0104-01 21-2120-0104-231 21-2120-0105-00 21-2120-0105-01 21-2120-0105-02L 21-2120-0105-03L 21-2120-0105-06L 21-2120-0105-07L 21-2120-0105-08 21-2120-0105-08L 21-2120-0105-12L 21-2120-0105-13L 21-2120-0105-14L 21-2120-0105-15L 21-2120-0105-17L 21-2120-0105-50 21-2120-0105-50L 21-2125-0104-01 21-2125-0105-03L 21-2125-0105-07L 21-2125-0105-08L 21-2125-0105-50L 21-2127-0104-01 21-2127-0105-01 21-2127-0105-02L 21-2127-0105-03L 21-2127-0105-06L 21-2127-0105-08L 21-2127-0105-50L

Distribution

Worldwide - US nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

185,975 US; 127,701 OUS Total

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1668-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.