Class II

Medical Device Recall: Medex 700PSI STOPCOCK W/ROTATOR, List Number MX4331R

Smiths Medical Asd Inc · November 10, 2023

Reason for Recall

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, UAE.

States Affected

NATIONWIDE

Quantity Affected

31,685 total

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0742-2024