Class II
Medical Device Recall: medex TranStar KIDS KIT 19IN. CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9543
Smiths Medical ASD, Inc. · September 24, 2024
Reason for Recall
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Distribution
Worldwide distribution.
Quantity Affected
800 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0197-2025
Status: ongoing