Class II

Medical Device Recall: Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500

Smiths Medical ASD Inc. · December 19, 2023

Reason for Recall

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

Product Description

Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500-402 j) 3500-414 k) 3500-415 l) 3500-500 m) 3500E n) 3500VX-306 o) 3500VX-500

Distribution

Worldwide distribution.

Quantity Affected

3821 pumps

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1540-2024

Status: ongoing