Class I

Medical Device Recall: smiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes, labeled as follows: 1) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilato

Smiths Medical ASD Inc. · June 7, 2024

Reason for Recall

There is the potential for a disconnection of the pilot balloon from the tracheostomy inflation line within specific lots of the BLUSelect, BLUgriggs and BLUperc products due to a manufacturing defect.

Product Description

smiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes, labeled as follows: 1) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 7.0 mm BLUselect trach tube, Item Number 101/561/070; 2) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 8.0 mm BLUselect trach tube, Item Number 101/561/080; 3) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 9.0 mm BLUselect trach tube, Item Number 101/561/090; 4) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 7.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/070; 5) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 8.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/080; 6) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilator and 9.0 mm BLUselect SuctionAid trach tube, Item Number 101/563/090; 7) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 7.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/070; 8) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 8.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/080; 9) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 9.0 mm BLUselect trach tube (drugs, US ONLY), Item Number 101/595/090; 10) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 7.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/070; 11) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 8.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/080; 12) BLUperc Percutaneous Dilation Procedural Tray with Single Stage Dilator and 9.0 mm BLUselect SuctionAid trach tube (drugs, US ONLY), Item Number 101/596/090

Distribution

Worldwide

Quantity Affected

41691 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3088-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

Smiths Medical ASD Inc. Medical Device Recall: smiths medical portex BLUperc Percutaneous Diliation Procedural Trays with BLUselect tracheostomy tubes, labeled as follows: 1) BLUperc Percutaneous Dilation Procedural Kit with Single Stage Dilato | SafeCheck