Medical Device Recall: smiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JP
Smiths Medical ASD, Inc. · October 28, 2024
Reason for Recall
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
Distribution
Worldwide distribution.
Quantity Affected
142327 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0527-2025
Status: ongoing