Class II

Medical Device Recall: Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to DVI-I male/male d

Spacelabs Healthcare, Inc. · September 5, 2025

Reason for Recall

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

Product Description

Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-01 Software Version: N/A Product Description: 6 foot DVI-I to DVI-I male/male display cable. Component: Yes, the cable is an external display cable used with the 91390 and 91393

Distribution

Worldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BGD, BHR, BOL, BRA, BRN, CAN, CHE, CHL, CHN, COL, CRI, CYP, CZE, DOM, DZA, ECU, EGY, ESP, FRA, GBR, GRC, GTM, GUM, HKG, IDN, IND, IRL, IRN, ISR, ITA, JOR, KWT, LBN, MAR, MEX, MYS, NLD, NZL, OMN, PAK, PAN, PER, PHL, PLS, POL, PRI, PRT, PRY, QAT, ROM, SAU, SGP, SLV, SRB, THA, TTO, TUN, TUR, TWN, VNM.

States Affected

NATIONWIDE

Quantity Affected

8179

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0167-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.