Medical Device Recall: Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor

Spacelabs Healthcare, Inc. · February 27, 2026

Reason for Recall

Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.

Distribution

Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.

States Affected

NATIONWIDE

Quantity Affected

1790

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1793-2026

Status: ongoing