Medical Device Recall: AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of s
Steris Corporation · November 19, 2025
Reason for Recall
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
Product Description
AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
Distribution
Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.
States Affected
NATIONWIDE
Quantity Affected
22
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0910-2026
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.