Class II
Medical Device Recall: Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems
Steris Corporation · June 23, 2025
Reason for Recall
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
Distribution
Worldwide - US Nationwide and Puerto Rico and the country of Canada.
States Affected
NATIONWIDE
Quantity Affected
167 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2177-2025
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.