Class II

Medical Device Recall: Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge,

Tandem Diabetes Care, Inc. · January 30, 2026

Reason for Recall

The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.

Product Description

Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).

Distribution

International distribution in the countries of Czech Republic, Slovakia.

States Affected

IN

Quantity Affected

1,380 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1636-2026

Status: ongoing