Class II

Medical Device Recall: TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

TAS Medical Inc · January 27, 2025

Reason for Recall

Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.

Distribution

US Nationwide distribution in the states of LA, OK, TN.

States Affected

NATIONWIDE

Quantity Affected

114

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1294-2025

Status: ongoing