Medical Device Recall: Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qua
The Binding Site Group, Ltd. · April 16, 2026
Reason for Recall
It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.
Product Description
Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
1
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2362-2026
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.