Class II

Medical Device Recall: TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

Tobii Dynavox Llc · July 8, 2026

Reason for Recall

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2959-2026

Status: ongoing