Class II
Medical Device Recall: TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Tobii Dynavox Llc · July 8, 2026
Reason for Recall
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2959-2026
Status: ongoing