Medical Device Recall: The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical informatio
Topcon Medical Systems, Inc. · September 17, 2024
Reason for Recall
The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided
Product Description
The IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices. It is intended for processing and displaying ophthalmic images and optical coherence tomography data. The IMAGEnet 6 Ophthalmic Data System uses the same algorithms and reference databases from the original data capture device as a quantitative tool for the comparison of posterior ocular measurements to a database of known normal subjects.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
3144 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0362-2025
Status: ongoing