Class II

Medical Device Recall: TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318

Tornier, Inc · December 19, 2025

Reason for Recall

A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

427

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1237-2026