Medical Device Recall: Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPola
Total Joint Orthopedics, Inc. · September 23, 2025
Reason for Recall
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Product Description
Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 3205.42.022 Klassic BiPolar Head, 42 mm OD, 22 mm Head; 3205.43.022 Klassic BiPolar Head, 43 mm OD, 22 mm Head; 3205.44.028 Klassic BiPolar Head, 44 mm OD, 28 mm Head; 3205.45.028 Klassic BiPolar Head, 45 mm OD, 28 mm Head; 3205.46.028 Klassic BiPolar Head, 46 mm OD, 28 mm Head; 3205.47.028 Klassic BiPolar Head, 47 mm OD, 28 mm Head; 3205.48.028 Klassic BiPolar Head, 48 mm OD, 28 mm Head; 3205.49.028 Klassic BiPolar Head, 49 mm OD, 28 mm Head; 3205.50.028 Klassic BiPolar Head, 50 mm OD, 28 mm Head; 3205.51.028 Klassic BiPolar Head, 51 mm OD, 28 mm Head; 3205.52.028 Klassic BiPolar Head, 52 mm OD, 28 mm Head; 3205.53.028 Klassic BiPolar Head, 53 mm OD, 28 mm Head; 3205.54.028 Klassic BiPolar Head, 54 mm OD, 28 mm Head; 3205.55.028 Klassic BiPolar Head, 55 mm OD, 28 mm Head; 3205.56.028 Klassic BiPolar Head, 56 mm OD, 28 mm Head; 3205.58.028 Klassic BiPolar Head, 58 mm OD, 28 mm Head; 3205.60.028 Klassic BiPolar Head, 60 mm OD, 28 mm Head
Distribution
US Nationwide distribution in the states of Ohio, Texas, Florida, Georgia, and North Carolina.
States Affected
NATIONWIDE
Quantity Affected
103 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0432-2026
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.