Class II
Medical Device Recall: TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only
TriMed Inc. · March 19, 2024
Reason for Recall
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Distribution
US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.
States Affected
NATIONWIDE
Quantity Affected
120 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1792-2024
Status: ongoing