Class II

Medical Device Recall: Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

Vantive US Healthcare LLC · August 28, 2025

Reason for Recall

Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician's remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom, Chile, and Colombia.

States Affected

NATIONWIDE

Quantity Affected

1

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0409-2026

Status: ongoing