Medical Device Recall: Brand Name: NIO-A Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000603 Software Version: N/A Product Description: NIO devices are a spring-based, automatic intra-osseous acc
Waismed Ltd. · June 5, 2025
Reason for Recall
Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.
Product Description
Brand Name: NIO-A Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000603 Software Version: N/A Product Description: NIO devices are a spring-based, automatic intra-osseous access device supplied Sterile and single use. Indicated for Intra-osseous access to the Proximal humerus and Proximal Tibia in adult patients, in emergent situations. Component: N/A
Distribution
Worldwide distribution: US (nationwide) to state of: NC and OUS (international) countries of: Slovenia, Australia, Brazil, Estonia, South Africa, U. Arab Emirate, Greece, Panama, Singapore, Italy, Germany, Portugal, Spain, Ukraine, Lithuania, Chile, Poland, Germany, New Zealand, Montenegro, Switzerland, Argentina, Poland, Austria, Sweden, United Kingdom, Italy, France, Romania, Netherlands, Croatia, Lithuania, New Zealand, Japan, Cyprus, and Israel.
States Affected
NATIONWIDE
Quantity Affected
7056 units (US)
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2118-2025
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.