Medical Device Recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
Wipro GE Healthcare Private Ltd. · September 5, 2025
Reason for Recall
GE HealthCare has identified an issue affecting Carescape" Central Station (CSCS) with software version V3.0.5, which can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Distribution
Worldwide distribution.
Quantity Affected
173 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0213-2026
Status: ongoing