Class II

Medical Device Recall: ZAP-X Radiosurgery System, Model: 300150

Zap Surgical Systems · July 25, 2025

Reason for Recall

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

Distribution

Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.

States Affected

NATIONWIDE

Quantity Affected

17

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2639-2025

Status: ongoing