Medical Device Recall: Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509400 Software Version: N/A Product Description: Affixus 9 mm, Length 400 m
Zimmer, Inc. · December 2, 2025
Reason for Recall
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Product Description
Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509400 Software Version: N/A Product Description: Affixus 9 mm, Length 400 mm, Right, Trochanteric Nail Component: N/A
Distribution
US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
States Affected
NATIONWIDE
Quantity Affected
23
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0971-2026
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.