Medical Device Recall: Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to a patient under the direction or supervision of physician or o

Zyno Medical LLC · December 20, 2023

Reason for Recall

When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death

Product Description

Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to a patient under the direction or supervision of physician or other certified health care professional Model: Z-800WF

Distribution

Nationwide

States Affected

NATIONWIDE

Quantity Affected

574 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1183-2024

Status: ongoing