Abbott Recalls
9 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (9)
- Class I
September 10, 2025
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, F, (UniD Curve F), REF A-TFSE-F; cardiac catheter
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
- Class I
September 10, 2025
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DF, (BiD Curve D-F), REF A-TFSE-DF; cardiac catheter
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
- Class I
September 10, 2025
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
- Class I
September 10, 2025
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheter
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
- Class I
September 10, 2025
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DD, (BiD Curve D-D), REF A-TFSE-DD; cardiac catheter
Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.
- Class II
March 17, 2025
Amplatzer TorqVue LP Delivery System (TVLP), Model nos. 9-TVLP4F90/060 and 9-TVLP4F90/080
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
- Class II
March 17, 2025
Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
- Class II
March 4, 2024
Alinity i Anti-HCV Reagent Kit, List Number 08P0521
Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).
- Class II
March 4, 2024
ARCHITECT Anti-HCV Reagent Kit, List Numbers: a) 1L79-25, b) 1L79-35
Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).