Abbott Diabetes Care, Inc. Recalls
6 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (6)
- Class I
November 24, 2025
FreeStyle Libre 3 Sensor REF 72081-01 UDI-DI code: 00357599818005 The FreeStyle Libre 3 sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- Class I
November 24, 2025
FreeStyle Libre 3 Plus Sensor REF 78768-01 UDI-DI code: 00357599844011 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- Class I
November 24, 2025
FreeStyle Libre 3 Plus Sensor REF 78769-01 UDI-DI code: 00357599843014 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- Class I
November 24, 2025
FreeStyle Libre 3 Plus Sensor REF 72080-01 UDI-DI code: 00357599819002 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System.
Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results
- Class I
July 24, 2024
FreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose Monitoring System
Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death.
- Class II
February 9, 2023
FreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 Operating System
If using affected glucose monitoring app on Android 13 Operating System, extended periods of signal loss may be experienced, due to app not connected, which could impact ability to receive glucose reading/alarms, which could lead to undetected low or high glucose, which could result in delayed treatment: not taking insulin (for high glucose), or not taking glucose (for low glucose) when required.