Abbott Molecular, Inc. Recalls
9 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (9)
- Class II
April 28, 2025
Alinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and diff
There were reports of an increase in reactive negative controls and false positive results.
- Class II
April 28, 2025
Alinity m Resp-4-Plex Amplification (AMP) Kit (CE), List Number 09N79-090, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the qualitative detection
There were reports of an increase in reactive negative controls and false positive results.
- Class II
March 20, 2025
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
- Class II
September 12, 2024
Abbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-091
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.
- Class II
September 12, 2024
Abbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REF 09N17-090; c) REF 09N17-091;
Abbott has identified an increase of incidences regarding Error Code (EC) 9198 (Positive control is non-reactive) while using the Alinity m HR HPV AMP Kit and Alinity m STI AMP Kit. Certain invalidated positive assay controls can be traced to iron leaching into the Alinity m Lysis Solution from the lysis transfer pump in the Alinity m System.
- Class II
March 20, 2024
Alinity m System, Part Number: 08N53-002
The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
- Class III
February 1, 2024
Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)
Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.
- Class III
February 1, 2024
In-vitro diagnostic kits: (1) Abbott Vysis LSI ATM (11q22.3) SpectrmOrange Probe, REF 01N33-020; (2) Abbott Vysis PTEN/CEP 10 FISH Probe Kit, REF 04N62-020; (3) Abbott Vysis TelVysion 1p Spectrum
Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.
- Class III
February 1, 2024
In-vitro diagnostic kits: (1) Abbott Vysis MDM2 / CEP 12 FISH Probe Kit, REF 01N15-010; (2) Abbott Vysis LSI D13S319 (13q14.3) SpectrumOrange Probe, REF 01N34-020;
Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.