Ambu Inc. Recalls
4 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (4)
- Class I
July 9, 2025
Ambu SPUR II Labeled as the following: 1 SPUR II Adult Resuscitator, Adult, Catalog Number: 523211000 2 SPUR II Adult Resuscitator w/Mercury Filter, Adult, Catalog Number: 523611051 3 SPUR II Adult Re
Potential for the manometer port being blocked rendering the manometer non-functional.
- Class II
April 7, 2025
Ambu aScope" 5 Broncho HD 5.6/2.8 Sampler Set. Model Number: 622002000US
Incorrect labeling in which the front red pouch label did not match the actual size of the medical device. The label on the front of the scope incorrectly states 5.0/2.2 while the back label, shipping box and product inside correctly states 5.6/2.8.
- Class II
September 16, 2024
Ambu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000)
Ambu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.
- Class II
December 5, 2023
Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2
Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.