Daavlin Distributing Company Recalls
4 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (4)
- Class II
March 5, 2026
Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Class II
March 5, 2026
Photera 600 3-D, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Class II
March 5, 2026
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Class II
December 10, 2025
Phothera 200, Phototherapy Devices indicated for use to treat diagnosed skin disorders such as, but not limited to, Psoriasis, Vitiligo, and Atopic Dermatitis (Eczema) under the direction of a physici
Some of the Calibration/Output Certificates had the outputs reversed.