Dentsply IH, Inc. Recalls
2 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (2)
- Class II
July 24, 2025
Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
- Class II
July 24, 2025
Brand Name: Atlantis IO FLO-S Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04B / 68020033 Software Version: n/a Product Description: The Atlantis intraoral feature locating object fo
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.