- Class II
January 31, 2025
DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement
Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
- Class II
January 15, 2025
DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement
The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.
- Class II
October 3, 2024
DeRoyal CHILDREN'S ORTHOPEDIC TRACEPACK, 89-9349-06
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
- Class II
October 3, 2024
DeRoyal CHILDREN'S GENERAL TRACEPACK, REF 89-9342-06
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
- Class II
October 3, 2024
DeRoyal CHILDREN'S LAPTRACEPACK, 89-9362.06
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
- Class II
October 3, 2024
DeRoyal C-SECTION TRACEPACK, REF 89-9338-06
DeRoyal is recalling certain lots of surgical procedure pack products that contain a Stryker SafeAir Smoke Evacuation Pencil. The recall is due to Stryker receiving reports that the pencil is activating without manual input when plugged into a power source or remains active after buttons are released.
- Class I
February 13, 2024
GeoMed Custom Tracecarts, REF Numbers: a) 53-1831 ORTHO TOTAL JOINT TRACECART, b) 53-1836 GENERAL SURGERY TRACECART
DeRoyal manufactured and distributed Surgical Tracecarts that contain a Nurse Assist 16FR Silver Urine Meter Foley which were subsequently recalled.
- Class II
November 22, 2023
DeRoyal CARDIAC CATH SET-UP KIT PGYBK, REF 89-10572
Kit were manufactured and distributed with Edwards Life Sciences TruWave Disposable pressure transducers which were subsequently recalled by Edwards Life Sciences.