Erbe USA Inc Recalls
4 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (4)
- Class II
June 26, 2026
ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000
A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.
- Class I
February 12, 2026
Flexible Cryoprobe (OD 1.7mm, L1.15mm) REF: 20402-410. For surgical use
Probes may rupture/burst during activation
- Class I
February 12, 2026
Flexible Cryoprobe (OD 1.1mm, L1.15mm) w/ oversheath (OD2.6mm, L817mm) REF: 20402-401 STERILE EO. For surgical use
Probes may rupture/burst during activation
- Class I
February 12, 2026
Flexible Cryoprobe (OD 2.4mm, L1.15mm) REF: 20402-411. For surgical use
Probes may rupture/burst during activation