- Class II
August 15, 2025
STRATAFIX Spiral PDS Plus Bidirectional Knotless Tissue Control Device, SFX SPI PDS+ BI VIO 14X14IN(36X36CM) USP1(M4) D/A MO-4. Product Code: SXPP2B400-11.
Potential for barb non-engagement.
- Class II
December 20, 2024
PERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic an
Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
- Class II
December 20, 2024
PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophtha
Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
- Class II
December 20, 2024
ETHIBOND EXCEL Polyester Suture, REF: 6664H, X425H, X872H, EH7491H. Not sold in US
Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
- Class II
October 8, 2024
MONOCRYL" (poliglecaprone 25) Suture REF Y936H MONOCRYL" Sutures are indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular and neurological tissu
Due a component mix-up in manufacturing, the affected products may contain incorrect needle type/size and suture length.
- Class II
September 25, 2024
Coated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358H
Complaints were received regarding needle pull-off and suture degradation. Analysis on returned product from this lot confirmed that some sutures were visually degraded in the package.
- Class II
May 10, 2024
VICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological tissue
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
- Class II
May 10, 2024
MONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated PRODUCT CODE: Y315H Y345H Y359H
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
- Class II
May 10, 2024
MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated. PRODUCT CODE: MCP266H13 M
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
- Class II
May 10, 2024
PDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in ophthalmic surgery. The
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
- Class II
May 10, 2024
PDS II (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in ophthalmic surgery. These sutures are p
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
- Class II
May 10, 2024
VICRYL Plus Antibacterial (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation, except for ophthalmic, cardiovascular and neurological tissues. PRODUCT CO
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
- Class II
April 23, 2024
SURGIFOAM" Absorbable Gelatin Powder Kit, Product Code 1979.
Some applicator tips from these lots had open seals.
- Class II
March 8, 2024
STRATAFIX Spiral MONOCRYL Plus Bidirectional Knotless Tissue Control Device: STRATAFIX SPIRAL MONOCRYL PLUS BI UD 12+12IN(30+30CM) USP3-0(M2) D/A PS PRM- Indicated for use in soft tissue approximatio
Degraded and unusable upon removal from the foil pouch due to exposed environmental conditions during transit and storage, compromising sterility