- Class II
September 11, 2025
ECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection, and/or creation of anastomoses, Product Number GLC80
The devices were shipped unsterilized.
- Class I
April 22, 2025
Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35
Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.
- Class II
March 7, 2025
PDS¿PLUS¿VIO¿30IN(75CM)¿USP6-0(M0.7)¿D/A¿C-1¿MP
Identified curing issues with the silicone during the needles manufacturing process.
- Class II
March 7, 2025
ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
Identified curing issues with the silicone during the needles manufacturing process.
- Class II
March 7, 2025
PDSII¿VIO¿30IN(75CM)¿USP6-0(M0.7)¿D/A¿C-1
Identified curing issues with the silicone during the needles manufacturing process.
- Class II
March 7, 2025
PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM
Identified curing issues with the silicone during the needles manufacturing process.
- Class II
December 11, 2023
ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping Surface technology), 45MM WHITE VASCULAR/THIN, Product No.
There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.