- Class II
May 2, 2024
GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW
GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.
- Class II
May 2, 2024
GE Vivid Ultrasound products, Model Numbers: a) Vivid E90, H45581LB; b) Vivid E95, H45581DA
GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.