ImaCor Inc. Recalls
2 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (2)
- Class II
August 29, 2025
ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
- Class II
September 16, 2024
ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010, CLT-010-1
Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.