Integra LifeSciences Corp. Recalls
47 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (47)
- Class II
February 6, 2025
AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's Product Numbers (Catalog Number): (1) ASX15/60 and (2) ASX15/80
Possibility for the obturator to break (separate).
- Class II
January 10, 2025
Brand Name: MediHoney¿ Gel Product Name: MediHoney¿ Wound & Burn Dressing Gel Model/Catalog Number: 31815 Software Version: N/A Product Description: MediHoney¿ Gel with Active Leptospermum Honey i
Potential that the induction seal is not completely sealed to the device tube packaging.
- Class II
December 16, 2024
CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device that provides constant intraventricular pressure and drainage of CSF for the management of
Potential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.
- Class II
December 16, 2024
CODMAN CERTAS Plus Programmable Valve Right Angle Valve Only. Model Number: 828820. Implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hyd
Potential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.
- Class II
December 16, 2024
CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810 Implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hyd
Potential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.
- Class II
September 26, 2024
Duraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694
Out of specification endotoxin levels.
- Class II
September 20, 2024
SURG PAT XRAY 1X3 Model/Catalog Number: 801408. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous sy
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous s
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 1/4X6 Model/Catalog Number: 801450. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous sys
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 3X3 Model/Catalog Number: 801409. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous s
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/4X1-1/2 Model/Catalog Number: 801397. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nerv
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/2X3 Model/Catalog Number: 801407. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central ner
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervo
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/4X1/4 Model/Catalog Number: 801399. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervo
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 1/2X6 Model/Catalog Number: 801451. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous sys
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 1/8X6 Model/Catalog Number: 801449. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous syst
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 2X6 Model/Catalog Number: 801455. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 3X6 Model/Catalog Number: 801456. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1X1 Model/Catalog Number: 801403. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous s
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/4X3 Model/Catalog Number: 801398. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 3-1/2X6 Model/Catalog Number:801457. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous sys
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 1X6-200 Model/Catalog Number: 801453. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous sy
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
MICR PATIE RND Model/Catalog Number: 801396. CODMAN Surgical Patties and CODMAN Surgical Strips are manufactured of COTTONOID¿ Material with x-ray detectable markers. All patties have a suture strin
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/2X1 Model/Catalog Number: 801402. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG PAT XRAY 1/2X1/2 Model/Catalog Number: 801400. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervo
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 20, 2024
SURG STRP 3/4X6 Model/Catalog Number: 801452. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous syst
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- Class II
September 13, 2024
MEDIHONEY Dressing with Active Leptospermum Honey Wound and Burn Dressing W/APPLICATOR, 1.5oz tube (Model Number 31515) and 3.5oz tube (Model Number 31535)
There is a potential for pinholes in the applicator pouch film which creates a sterility concern.
- Class II
July 9, 2024
TruDi NAV Suction, 0 Degrees; Model/Catalog Number: TDNS000Z. TruDi" NAV Suction Instruments are intended for use with the TruDi" Navigation System during surgical procedures in ENT medicine and s
Specific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on the system.
- Class II
June 7, 2024
AURORA Surgiscope System. Sterile, single use device that contains a Sheath, Obturator, and Imager.
Potential sheath damage that could lead to tissue damage and/or track hemorrhage.
- Class II
May 29, 2024
CUSA Excel C2600 23KHz Straight Handpiece-Indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft and hard tissue Catalog Number: C2600* *Note: you may s
Potential for the CUSA Excel 23KHz Straight Handpiece housing to crack, may result in delay of treatment
- Class II
May 7, 2024
CODMAN EVD BACTISEAL 1.9mm I.D. Catheter Set- indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) a
Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.
- Class II
May 7, 2024
CODMAN BACTISEAL EVD Catheter Set - Indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other
Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal that may compromise the product sterility.
- Class II
May 7, 2024
CODMAN BACTISEAL Clear EVD Catheter 1.9mm I.D. Catheter Set- indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospina
Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.
- Class II
April 17, 2024
Integra¿ Cranial access kit (without Prep solutions) Bit and Guard- Used during a ventriculostomy procedure. Catalog Number: INS5HND
Potential holes and tears in the sterile packaging (header bag) of the Cranial Access Kit may compromise the sterility of the device.
- Class II
January 19, 2024
Ref No: INSHITH / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: INS7260 / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: 31156 / COMBO KIT CONTAINS INS-8301, INS-4500,INS-HITH. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: INS7270 / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: INS7250 / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: INSHITHRZN / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: INS5HND / Integra Cranial access kit (without Prep solutions) Bit and Guard. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
INS7280 / Integra¿ Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: HITHSP06 / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: INSHITHND / Integra¿ Cranial access kit (without Prep solutions) Bit and Guard. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.
- Class II
January 19, 2024
Ref No: HITHSP04 / Integra Cranial Access Kit. For access to the subarachnoid space or the lateral ventricles of the brain.
A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised.