Johnson & Johnson Surgical Vision, Inc. Recalls
2 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (2)
- Class II
March 14, 2024
Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the sh
Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
- Class II
December 12, 2023
TECNIS Toric II OptiBlue IOL Models ZCW
Due to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle between Low Power Meridian (LPM) and toric cylinder axis marks.
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.