Medtronic Neurosurgery Recalls
8 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (8)
- Class II
June 17, 2026
Ventriculostomy Kit, REF: 46154
Ventriculostomy devices, used during ventriculostomy procedures to gain access to the ventricular system, may exhibit endotoxin levels that exceed the acceptable limit for devices that come into contact with cerebrospinal fluid; endotoxin exposure can cause an acute inflammatory process.
- Class I
November 1, 2024
Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
- Class I
November 1, 2024
Exacta External Drainage and Monitoring System: 27581 KIT 27581 EXACTA W/EDM VCATH 35CM 27636 EDM 27636 EXACTA W/BRAIDED CORD LOCK 27666 EDM 27666 EXCTA W GLUE STPCK GR PT LN 27732 EDM 27732 EXACT
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
- Class I
January 22, 2024
Medtronic Duet External Drainage and Monitoring System, SmartSite, Injection Sites, REF 46914, Sterile EO, Rx Only. Used to drain cerebrospinal fluid (CSF) as a means to reduce intracranial pressure
Potential for catheter disconnection from the patient line stopcock connectors.
- Class I
January 22, 2024
Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, REF 46913, Sterile EO, Rx Only. Used to drain cerebrospinal fluid (CSF) as a means to reduce intracranial pressure a
Potential for catheter disconnection from the patient line stopcock connectors.
- Class I
January 22, 2024
Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Lumbar Catheter, REF 46917, Sterile EO, Rx Only. Used to drain cerebrospinal fluid (CSF) as a means to reduce intracr
Potential for catheter disconnection from the patient line stopcock connectors.
- Class I
January 22, 2024
Medtronic Duet External Drainage and Monitoring System, SmartSite Injection Sites, Ventricular Catheter, REF 46916, Sterile EO, Rx Only. Used to drain cerebrospinal fluid (CSF) as a means to reduce in
Potential for catheter disconnection from the patient line stopcock connectors.
- Class I
January 22, 2024
Medtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Ventricular Catheter, REF 46915, Sterile EO, Rx Only. Used to drain cerebrospinal fluid (CSF) as a means to reduce in
Potential for catheter disconnection from the patient line stopcock connectors.
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.