PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Recalls
100 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (100)
- Class II
July 16, 2026
Philips Azurion Rystem, Model Numbers: 722063, 722064, 722067, 722068, 722078, 722079, 722221, 722222,722223, 722224, 722225, 722226, 722227, 722228, 722280*, 722281*, and 722282* (*These models ar
Two software related issues affecting Azurion R2.2.10 Systems: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
- Class II
July 16, 2026
Philips Azurion, Model Numbers: 722229, 722230, 722231, 722232, 722233, 722234, 722235, and 722236, with Software Version Numbers: R3.0 and R3.1; Interventional fluoroscopic x-ray system
Two software related issues: 1) Potential image quality degradation that may lead to delay of therapy, procedural complications and/or inappropriate treatment and 2) Potential loss of geometry movements that may lead to delay of therapy and procedural complications
- Class II
June 29, 2026
Allura Xper FD. Model Numbers: 722026, 722027, 722028, 722029, 722035, 722038, 722039, 722058. Interventional fluoroscopic x-ray system.
Firmware issue potentially leading to loss of geometry movements.
- Class II
June 17, 2026
Philips Allura, Cardio Vascular-Allura Centron. Model Number: 722400. Interventional Fluoroscopic X-Ray System.
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
- Class II
June 17, 2026
Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280.
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
- Class II
June 17, 2026
Philips Allura under the following System Descriptions and corresponding model numbers: 1. Allura Xper FD10; Model Numbers: 722003, 722010, 722026. 2. Allura Xper FD10 OR Table; Model Number: 7220
Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.
- Class II
May 18, 2026
Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost E
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
- Class II
May 18, 2026
Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
- Class II
April 30, 2026
Allura system; System Code Description (Model Numbers): Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10 OR Table (722022, 722033), Allura Xper FD10/10 (722005, 722011, 722027), Allura Xper
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
- Class II
April 30, 2026
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20 (722228), Azurion 7 B12 (722067, 722225),
It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.
- Class II
April 28, 2026
Allura system; System Code Description (Model Numbers): Allura Xper FD10C (722001), Allura Xper FD10F (722002), Allura Xper FD10 (722003, 722010, 722026), Allura Xper FD10/10 (722005, 722011, 722027),
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
- Class II
April 28, 2026
Integris-Allura system; System Code Description (Model Numbers): Integris CV Cesar-Powerpack-Visub-Nicol (722030), Integris Allura 15 & 12 (monoplane) (722043), INTEGRIS Allura 15-12 (biplane) (722044
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
- Class II
April 28, 2026
Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding model numbers. 1. Azurion 3 M12; Model Numbers: 722063, 7222
Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.
- Class II
April 28, 2026
Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079,
The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.
- Class II
March 31, 2026
Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 7222
Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
- Class II
March 19, 2026
Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Philips interventional X-ray system and 3DRA d
Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.
- Class II
March 5, 2026
Vue Motion V12. Product Number: 1017979.
Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence.
- Class II
March 3, 2026
Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Azurion 5 M12; System Code: (1)722227, (2)722231;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Allura Xper FD20/20; System Code: 722038;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Allura Xper FD20/15; System Code: 722058;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Allura Xper FD10C; System Code: 722001;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Allura Xper FD20/20 OR Table; System Code: 722039;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD10F; System Code: 722002;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Allura Xper FD20/15 OR Table; System Code: 722059;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD20; System Code: (1) 722006, (2)722012, (3)722028;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
March 3, 2026
Allura Xper FD20/10; System Code: 722029;
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
- Class II
January 26, 2026
Azurion 3 M15 System Model Numbers: (1) 722064, (2) 722222, (3) 722230, (4)722280(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
January 26, 2026
Azurion 5 M20 System Model Numbers: (1)722228, (2)722232, (3)722281(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
January 26, 2026
Azurion 7 M12 System Model Numbers: (1)722078, (2)722223, (3)722233; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
January 26, 2026
Azurion 7 B12 System Model Numbers:(1) 722067, (2) 722225, (3) 722235; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
January 26, 2026
Azurion 7 M20 System Model Numbers: (1)722079, (2)722224, (3)722234, (4)722282(OUS only); Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
January 26, 2026
Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
January 26, 2026
Azurion 3 M12 System Model Numbers: (1) 722063, (2) 722221, (3) 722229; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
January 26, 2026
Azurion 7 B20 System Model Numbers: (1)722068, (2)722226, (3)722236; Software Version Number - All versions: R1.X, R2.X, R3.X
Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.
- Class II
December 15, 2025
Allura Xper FD20; Model Number: 722012;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD10F; Model Number: 722002;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD10/10; Model Number: 722027;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD20; Model Number: 722006;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD10; Model Number: 722026;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Product Name: Allura Xper FD10C; Model number: 722001;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD10; Model Number: 722003;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD20 Biplane; Model Number: 722008;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD20 OR Table; Model Number: 722023;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD20 Biplane; Model Number: 722013;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD10/10; Model Number: 722011;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD20; Model Number: 722028;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD20/20; Model Number: 722038;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD10/10; Model Number: 722005;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD20/10; Model Number: 722029;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 15, 2025
Allura Xper FD10; Model Number: 722010;
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Class II
December 12, 2025
Philips Azurion system; Software Version Number: R3.1;
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
- Class II
December 12, 2025
Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azuri
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
- Class II
December 8, 2025
Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion
Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
- Class II
October 2, 2025
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
- Class II
September 29, 2025
Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
- Class II
September 3, 2025
Allura Xper FD20/20; Model Numbers: 722038; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD10C; Model Number: 722001; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20/20 OR Table; Product Code: 722039; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper CV20; Model Numbers: 722031; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD10/10 Product Codes: (1)722011, (2) 722027, (3) 722005; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20 Product Codes: (1)722012, (2) 722028, (3) 722006; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD10 OR Table; Model Numbers: 722033; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Cardio Vascular-Allura Centron; Model Numbers: 722400; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20 Biplane OR Table; Model Numbers: (1) 722020, (2) 722025; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD20/10; Product Code: 722029; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD10 OR Table Product Codes: (1)722022, (2) 722033; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD20 Biplane; Model Numbers: (1) 722008, (2) 722013; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD10 Product Codes: (1)722010, (2) 722026, (3) 722003; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20 OR Table Product Codes: (1)722023, (2) 722035, (3) 722015; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;
Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Class II
September 3, 2025
Allura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD20; Model Numbers: (1) 722006, (2) 722012, (3) 722028; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
September 3, 2025
Allura Xper FD10/10; Model Numbers: (1) 722005, (2) 722011, (3) 722027; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
- Class II
August 8, 2025
IntelliSpace Cardiovascular, Software 8.0.0.4.
Software issue that results in the display of outdated information.
- Class II
August 1, 2025
Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
- Class II
August 1, 2025
Philips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003 2. Allur
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
- Class II
August 1, 2025
Interventional Fluoroscopic X-Ray System - Philips Allura Xper Series. Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 722003. 2. Allura Xper FD10/1
Potential that a component (air baffle) may be missing in some Nehalem FlexVision-2 PCs that are part of the Allura Xper series.