Qiagen Sciences LLC Recalls
3 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (3)
- Class II
March 12, 2025
Brand Name: QIAstat-Dx Product Name: QIAstat-Dx Respiratory Panel Plus Reference Number (REF): 691224 Software Version: n/a Component: n/a
Identified faulty cartridges. If such a faulty cartridge is used for sample testing, false test results could occur.
- Class II
June 3, 2024
QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model/Catalog Number: 900
Heater shaker module does not perform heating if the temperature is set below 40¿C for static incubation steps when running IVD protocol QIAamp¿ DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA from Gram-positive bacteria (Script filename: Kt_name_Bacteria (Gram+) or rY'east_Enzymatic Lysis_ V2) which could potentially cause delayed or erroneous results depending on the downstream assay.
- Class II
March 4, 2024
EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model Number: 9003230
Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost