Remel Europe Ltd. Recalls
2 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (2)
- Class II
July 17, 2026
Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics
Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.
- Class II
March 25, 2026
Oxoid Agglutinating Sera, Salmonella 9-0 R30957301
Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.