Sophysa S.A. Recalls
2 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (2)
- Class II
April 9, 2024
The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure and temperature with tunneling (PSO-VTT) indicated for use
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
- Class II
April 9, 2024
The Sophysa Pressio¿ Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricu
An increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.