UIH Technologies LLC Recalls
9 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (9)
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 760 MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uEXPLORER MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT ATLAS MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI Panorama 35 MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 780 MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
- Class II
September 12, 2025
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 780 MD, Rx Only
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.