Medical Device Recalls (5)
- Class II
August 26, 2025
GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A,
Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.
- Class II
July 3, 2024
GORE SEAMGUARD Bioabsorbable Staple Line Reinforcement Configured for Circular Staplers, 25mm; Catalog number: 1BSGC25;
Devices were labeled with an expiration date of four years and ten days rather than the validated two years.
- Class II
July 1, 2024
Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB051501A BXB051502A BXB051901A BXB051902A BXB052901A BXB052902A BXB053901A BXB0
Due to packaging defects, sterility assurance and heparin activity may be compromised.
- Class II
July 1, 2024
Heparin, Gore Viabahn VBX, Balloon Expandable Endoprosthesis, for the following Reference numbers: United States: BXB055902C, BXB063902C, BXB065902C, BXB067902C, BXB073902C, BXB075902C, BXB077902C
Due to packaging defects, sterility assurance and heparin activity may be compromised.
- Class II
January 19, 2024
GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with holes; and (2) Catalog Number VSWVN1008, 10 mm x 8 cm; no holes
Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary procedure, cholangitis, cholecystitis, and/or pain. The GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis is a flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. The endoprosthesis is radially compressed and secured onto the distal end of a catheter-based delivery system for deployment. The catheter provides a means for accessing and implanting the GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis at the target site in the biliary tract.