ZOLL Medical Corporation Recalls
10 recalls from federal safety databases (FDA, CPSC). Sourced directly from government enforcement feeds.
Medical Device Recalls (10)
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO. Software Version: NA Product D
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (A
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80A; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for t
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C; Software Version: NA Product Description: The Powerheart G5 automated ext
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-01C, (2) G5A-02C, (3) G5A-05C, (4) G5A-08C, (5) G5A-10C, (6) G5A-13C, (7) G5A-15C, (8) G5A-22C, (9) G5A-41C, (10) G5A-8
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for t
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-00A-TSO, (2) G5S-01A, (3) G5S-02A, (4) G5S-05A, (5) G5S-08A, (6) G5S-10A, (7) G5S-11A, (8) G5S-80A-TSO, (9) G5S-80-L (1
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
February 13, 2025
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80C, (2) G5A-80C-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (A
The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Class II
May 31, 2024
ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, tha
G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest
- Class I
April 30, 2024
ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference Guide (QRGs) as follows: Part Numbers: 1. 799-EGL2-02 2. 799-EGL2-02-01 3. 799-EGL2
Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.